How to use IDEXX QC kits?
Step 1
Remove only the necessary vial(s) from freezer. Equilibrate at room temperature for 15 minutes.
Step 2
Open vial and aseptically transfer the pellet to an appropriately labeled vessel containing 100 mL of room temperature sterile water.
Step 3
Swirl the sample and allow to stand for up to 5 minutes. The pellet should completely dissolve. After it is dissolved, mix by inverting the vessel 10 times.
Step 4
Refer to the acceptance range for each organism on the Certificate of Analysis. The IDEXX-QC kit lot can be found on the box label.
A complete suite of microbiology QC kits
- Acceptance ranges that are calculated using IDEXX reagents.
- Source organisms are no more than two passages from primary cultures.
- Use IDEXX-QC for presence/absence or quantification.
- All materials are ISO 17034:2016 certified.
Product Information
IDEXX-QC Coliform and E. coli Kit
- Contains 3 E. coli, 3 K. variicola, 3 P. aeruginosa.
- For use with Colilert, Colilert-18, and Colisure tests.
- Catalog number: WQC-TCEC
IDEXX-QC Fecal Coliform Kit
- Contains 3 E. coli, 3 P. aeruginosa.
- For use with the Colilert-18 Test.
- Catalog number: WQC-FC
IDEXX-QC Enterococci Kit
- Contains 3 E. faecalis, 3 E.coli, 3 S. bovis.
- For use with Enterolert and Enterolert-E tests.
- Catalog number: WQC-ENT
IDEXX-QC Legionella pneumophila Kit
- Contains 3 L. pneumophila, 3 E. faecalis.
- For use with the Legiolert Test.
- Catalog number: WQC-LP
IDEXX-QC Pseudomonas aeruginosa Kit
- Contains 3 P. aeruginosa, 3 E. coli, 3 P. fluorescens.
- For use with the Pseudalert Test.
- Catalog number: WQC-PSE
IDEXX-QC Enterococci Drinking Water Kit
- Contains 3 E. faecalis, 3 S. marcescens.
- For use with the Enterolert-DW Test.
- Catalog number: WQC-ENTDW
IDEXX-QC HPC/TVC Kit
- Contains 3 E. faecalis.
- For use with EasyDisc tests, SimPlate for HPC, and HPC for Quanti-Tray.
- Catalog number: WQC-HPC
IDEXX-QC E. coli Kit
- Contains 9 E. coli.
- For use with Colilert, Colilert-18, and Colisure tests.
- Catalog number: WQC-EC